FDA Adverse Event Malfunction Summary report: N

ETS FLEX ARTICNG

MDR report key: 2173072 · Received July 23, 2011

Report

Report Number
3005075853-2011-02955
Event Type
Malfunction
Date Received
July 23, 2011
Report Date
June 28, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLECTOMY PROCEDURE, THE DEVICE WOULD NOT OPEN. IT IS UNKNOWN IF IT WAS ON TISSUE. NO OTHER DETAILS OF THE EVENT ARE AVAILABLE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETS FLEX ARTICNG STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK H43K5P

Patients

Seq Age Sex Outcome Treatment
1