FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2173068 · Received July 23, 2011

Report

Report Number
2531779-2011-05145
Event Type
Malfunction
Date Received
July 23, 2011
Report Date
June 24, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED AND TESTED WITH THE FOLLOWING FINDINGS: TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE KEYPAD. MULTIPLE BUTTON PRESSES ARE REQUIRED BEFORE THE BUTTONS WILL ENGAGE. PRODUCT ANALYSIS REMOVED KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS. CONTAMINATION WAS PRESENT UNDER THE CONTACTS OF ALL BUTTONS.

Description of Event or Problem · 1

REPORTER ALLEGED THAT THE KEYPAD IS WORN OUT AND ALL BUTTONS NEED TO BE PRESSED SEVERAL TIMES BEFORE THEY FUNCTIONED PROPERLY. MOTHER ALSO MENTIONED THAT THE SCREEN WAS DIM AND HAS NOTICED IT GRADUALLY GETTING DIMMER OVER TIME. THE KEYPAD IS REPORTEDLY INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 8 YR