ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 2173028
·
Received July 23, 2011
Report
- Report Number
- 2250051-2011-00173
- Event Type
- Malfunction
- Date Received
- July 23, 2011
- Date of Event
- July 16, 2011
- Report Date
- July 19, 2011
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- MMH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT. THE FSE FOUND THE TIP CLAMP AND PLUNGER CLAMP WERE BROKEN AND DIRTY WITH DRIED SERUM. THE FSE REPLACED THE PLUNGER CLAMP, TIP CLAMP AND LLD SPRING FOR COLLET #10 AND INSPECTED AND CLEANED ALL THE OTHER CLAMPS. THE FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.
Description of Event or Problem · 1
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE AND REAGENT AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).