ENDOPATH** ETS FLEX45
Report
- Report Number
- 3005075853-2011-02945
- Event Type
- Malfunction
- Date Received
- July 23, 2011
- Date of Event
- May 16, 2011
- Report Date
- May 24, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
(B)(4). FIRING TRIGGER TEETH. THE ANALYSIS RESULTS FOUND THAT THE ATW45 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED FULLY LOADED WITH STAPLES. NO FUNCTIONAL TEST COULD BE PERFORMED ON THE RETURNED DEVICE, HOWEVER, THE RELOAD WAS TESTED FOR FUNCTIONALITY WITH A TEST DEVICE. THE DEVICE FIRED, CUT, AND FORMED THE STAPLES AS INTENDED. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. WHILE NO CONCLUSION COULD BE REACHED AS TO WHAT CAUSED THE FIRING MECHANISM TO FAIL, IT IS POSSIBLE THAT THE DEVICE WAS FIRED ON TISSUE THICKER THAN INDICATED OR THROUGH A LOCKED CARTRIDGE IN PREVIOUS FIRINGS. WHEN EITHER OR BOTH OF THESE SCENARIOS OCCUR THE COMPONENTS IN THE FIRING MECHANISM CAN BE DAMAGED. IT SHOULD BE NOTED THAT A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4).
IT WAS REPORTED THAT DURING A LAP RECTUM PROCEDURE, THE INSTRUMENT DIDN'T FIRE. THE WHITE RELOAD PRESENT IN THE INSTRUMENT WAS CHANGED TO A NEW, GREEN RELOAD; HOWEVER, IT DIDN`T FIRE AS WELL. SURGERY WAS FINISHED WITH ANOTHER INSTRUMENT. SURGERY WAS PROLONGED TEN MINUTES. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH** ETS FLEX45 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | H43A5T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELOAD-TR45G |