FDA Adverse Event Malfunction Summary report: N

ENDOPATH** ETS FLEX45

MDR report key: 2172987 · Received July 23, 2011

Report

Report Number
3005075853-2011-02945
Event Type
Malfunction
Date Received
July 23, 2011
Date of Event
May 16, 2011
Report Date
May 24, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). FIRING TRIGGER TEETH. THE ANALYSIS RESULTS FOUND THAT THE ATW45 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED FULLY LOADED WITH STAPLES. NO FUNCTIONAL TEST COULD BE PERFORMED ON THE RETURNED DEVICE, HOWEVER, THE RELOAD WAS TESTED FOR FUNCTIONALITY WITH A TEST DEVICE. THE DEVICE FIRED, CUT, AND FORMED THE STAPLES AS INTENDED. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. WHILE NO CONCLUSION COULD BE REACHED AS TO WHAT CAUSED THE FIRING MECHANISM TO FAIL, IT IS POSSIBLE THAT THE DEVICE WAS FIRED ON TISSUE THICKER THAN INDICATED OR THROUGH A LOCKED CARTRIDGE IN PREVIOUS FIRINGS. WHEN EITHER OR BOTH OF THESE SCENARIOS OCCUR THE COMPONENTS IN THE FIRING MECHANISM CAN BE DAMAGED. IT SHOULD BE NOTED THAT A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP RECTUM PROCEDURE, THE INSTRUMENT DIDN'T FIRE. THE WHITE RELOAD PRESENT IN THE INSTRUMENT WAS CHANGED TO A NEW, GREEN RELOAD; HOWEVER, IT DIDN`T FIRE AS WELL. SURGERY WAS FINISHED WITH ANOTHER INSTRUMENT. SURGERY WAS PROLONGED TEN MINUTES. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** ETS FLEX45 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK H43A5T

Patients

Seq Age Sex Outcome Treatment
1 RELOAD-TR45G