FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 2172954 · Received July 23, 2011

Report

Report Number
2124215-2011-10868
Event Type
Malfunction
Date Received
July 23, 2011
Date of Event
April 24, 2011
Report Date
December 17, 2012
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS OF TODAY, NO ADDITIONAL INFORMATION IS AVAILABLE AND OUR INVESTIGATION IS COMPLETE AT THIS TIME. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. THE CLINICAL OBSERVATIONS WERE NOT ABLE TO BE CONFIRMED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT NOISE WAS SEEN ON THE RIGHT VENTRICULAR (RV) CHANNEL OF THIS IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) SYSTEM. THE NOISE WAS OVERSENSED WHICH LEAD TO PACING INHIBITION WITH GREATER THAN TWO SECONDS OF ASYSTOLE FOR THE PATIENT. TECHNICAL SERVICES DISCUSSED THE POSSIBLE CAUSES OF THE OBSERVED NOISE. THERE WERE NO ADVERSE PATIENT EFFECTS. THE SYSTEM REMAINS IN SERVICE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION AND RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T165

Patients

Seq Age Sex Outcome Treatment
1 66 YR T165| 4135| 0184