FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* TROCAR

MDR report key: 2172896 · Received July 22, 2011

Report

Report Number
3005075853-2011-02939
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
July 6, 2011
Report Date
July 6, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. UPON EVALUATION OF THE DEVICE, IT WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SLEEVE GASTRECTOMY PROCEDURE, TOWARDS THE END OF THE PROCEDURE THE TROCAR BEGAN TO LEAK. THE LEAK OCCURRED ONLY WHEN AN INSTRUMENT WAS INSERTED INTO IT. THE TROCAR WAS CONTINUED TO BE USED AND THE AIR LEAK WAS MANAGED. THE TROCAR DID NOT LEAK WHEN AN INSTRUMENT WAS NOT INSERTED. THERE WERE NO PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL* TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1