Report
- Report Number
- 6000001-2011-13519
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- July 1, 2011
- Report Date
- July 7, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE DEVICE WILL NOT BE RETURNED TO BAXTER BECAUSE IT WAS SERVICED ON-SITE AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION OF DAMAGED BATTERY WAS CONFIRMED BY BAXTER PERSONNEL DURING ON-SITE PRODUCT EVALUATION. THE ROOT CAUSE WAS ASSIGNED TO DEPLETED BATTERIES RESULTING FROM USE/USER ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. THIS EVENT OCCURRED UPON POWER UP; HOWEVER, IT IS UNKNOWN IN WHICH CARE AREA THIS EVENT OCCURRED. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |