SYNCHRON LX®20 PRO CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2050012-2011-03086
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 17, 2011
- Report Date
- June 17, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K011213
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
ERRONEOUS DATA GENERATED.
THE CUSTOMER CONTACTED BECKMAN COULTER, INC. (BCI) TO REPORT ERRONEOUSLY HIGH TRIGLYCERIDE (TG) RESULTS FOR TWO PATIENT SAMPLES ASSAYED WITH TRIGLYCERIDE REAGENT ON A SYNCHRON LX 20 PRO CLINICAL CHEMISTRY SYSTEM. THE HIGH TG RESULTS FOR THE TWO PATIENT SAMPLES WERE NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT. THE CUSTOMER RE-ASSAYED BOTH PATIENT SAMPLES FOR TG ON ANOTHER ANALYZER IN THEIR LABORATORY AND OBTAINED LOWER RESULTS WHICH WERE INTERPRETED TO BE CORRECT. THE CUSTOMER HAD RUN QUALITY CONTROLS FOR TG PRIOR TO THE EVENT. THE BIO RAD LEVEL 2 CONTROL HAD RECOVERED HIGH WITH RESPECT TO THE LABORATORY'S ESTABLISHED RANGE. THE CUSTOMER REPEATED THE TG ASSAY FOR THE BIO RAD LEVEL 2 CONTROL WHICH YIELDED AN ACCEPTABLE RESULT. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE. THE FSE OBSERVED THAT THE WASH TOWER ON THE ANALYZER WAS NOT FUNCTIONING PROPERLY. IN ADDITION, LEAKS WERE OBSERVED COMING FROM THE REAGENT PROBE A WASH COLLAR. THE FSE REPLACED THE WASH STATION AND THE REAGENT PROBE A WASH COLLAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON LX®20 PRO CLINICAL CHEMISTRY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | TRIGLYCERIDE REAGENT |