FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2172755 · Received July 22, 2011

Report

Report Number
2531779-2011-05127
Event Type
Malfunction
Date Received
July 22, 2011
Report Date
June 23, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE ISSUE WITH THE KEYPAD BUTTON MALFUNCTION WAS CONFIRMED THROUGH PRODUCT ANALYSIS. THE UP BUTTON IS INTERMITTENTLY RESPONDING TO USER INPUTS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE KEY CONTACTS.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE ALLEGED KEYPAD ISSUE. REPORTEDLY, THE UP AND DOWN BUTTONS HAS A DELAY RESPONSE. THE ISSUE WAS NOT RESOLVED WITH TRAINING. THERE WAS NO EVIDENCE OF PRODUCT MISUSE. THE ANIMAS PRODUCT IS BEING RETURNED FOR INVESTIGATION. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 60 YR