FDA Adverse Event Malfunction Summary report: N

OT ULTRALINK METER

MDR report key: 2172714 · Received July 22, 2011

Report

Report Number
2939301-2011-06103
Event Type
Malfunction
Date Received
July 22, 2011
Report Date
June 27, 2011
Manufacturer
LIFESCAN INC.
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE 510(K)# IS K073231. THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE FAILED TESTING. THE METER WAS FOUND TO HAVE PCB CONTAMINATION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE POWER BUTTON WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 3141365

Patients

Seq Age Sex Outcome Treatment
1