UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2172698
·
Received July 22, 2011
Report
- Report Number
- 2050012-2011-03311
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 18, 2011
- Report Date
- June 18, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 FOR THIS EVENT. THE FSE DISCOVERED THAT THE CONNECTOR GOING TO THE DRAIN TUBE TO THE VACUUM COLLECTOR WAS PLUGGED. THE FSE REPLACED THE CONNECTOR AND THIS RESOLVED THE ISSUE. BEC IDENTIFIER FOR THIS REPORT IS (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE CAP PIERCER IS LEAKING IN THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |