UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2172698 · Received July 22, 2011

Report

Report Number
2050012-2011-03311
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 18, 2011
Report Date
June 18, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 FOR THIS EVENT. THE FSE DISCOVERED THAT THE CONNECTOR GOING TO THE DRAIN TUBE TO THE VACUUM COLLECTOR WAS PLUGGED. THE FSE REPLACED THE CONNECTOR AND THIS RESOLVED THE ISSUE. BEC IDENTIFIER FOR THIS REPORT IS (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE CAP PIERCER IS LEAKING IN THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED IN THIS EVENT.