FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2172522 · Received July 22, 2011

Report

Report Number
6000001-2011-13473
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 22, 2011
Report Date
July 8, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A DISPLAY COVER TORN WAS CONFIRMED DURING SERVICE EVALUATION. THE QUALITY ENGINEER HAS ASSIGNED THE CAUSE TO A DAMAGED BEZEL. THE FRONT BEZEL WAS REPLACED TO CORRECT THE REPORTED CONDITION.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A TORN DISPLAY COVER. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THE EVENT OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1