FDA Adverse Event
Injury
Summary report: N
MATRIX12 M ESOPHAGEAL TEMPERATURE PROBE
MDR report key: 21724964
·
Received March 28, 2025
Report
- Report Number
- 3009437315-2025-00001
- Event Type
- Injury
- Date Received
- March 28, 2025
- Date of Event
- March 3, 2025
- Report Date
- March 28, 2025
- Manufacturer
- CIRCA SCIENTIFIC, INC.
- Product Code
- FLL
- UDI-DI
- 10851498007135
- PMA / PMN Number
- K222311
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THE RETURNED MATRIX12 M PROBE MET DIMENSIONAL SPECIFICATIONS AND NO DEFECTS OBSERVED THAT WOULD CAUSE DIFFICULT PLACEMENT. ROOT CAUSE IS INDETERMINATE, IT IS NOT KNOWN WHY USER EXPERIENCED DIFFICULT PLACEMENT.
Description of Event or Problem · 0
ANESTHESIA STATED DIFFICULT PLACEMENT WITH MATRIX12 M PROBE. PROBE THOUGHT TO BE IN ESOPHAGUS, DEPLOYED AND PLACEMENT FOUND TO BE IN BRONCHIAL STEM. BLEEDING OCCURRED. NOTHING UNUSUAL ABOUT ANATOMY ACCORDING TO ANESTHESIA & STAFF. THE LAB ALSO STOCKS S-CATH M. THIS WAS USED AFTER MATRIX12 M WAS REMOVED. THEY EXTUBATED SUCCESSFULLY AFTER THE CASE WAS FINISHED USING S-CATH M.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2624785 | MATRIX12 M ESOPHAGEAL TEMPERATURE PROBE | ESOPHAGEAL TEMPERATURE PROBE | FLL | CIRCA SCIENTIFIC, INC. | CS-31EP | 14659 | 10851498007135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |