FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2172467 · Received June 17, 2011

Report

Report Number
2032227-2011-01484
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
June 2, 2011
Report Date
June 2, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE CUSTOMER STATED THAT THE DEVICE WAS EXPOSED TO MAGNETIC RESONANCE IMAGING. TROUBLESHOOTING WAS PERFORMED. RAN A DISPLACEMENT TEST AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-522LNAH

Patients

Seq Age Sex Outcome Treatment
1 45 YR