FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 2172446
·
Received June 17, 2011
Report
- Report Number
- 1824206-2011-03312
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- May 31, 2011
- Report Date
- May 31, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACCOUNT REPLACED THE BRIDGE RECTIFIER TO REPAIR THE BED.
Description of Event or Problem · 1
THE ACCOUNT STATED THEY HAVE A BED IN STORAGE AND FOUND THE BED HAS NO POWER. WHEN HE PLUGGED THE BED INTO AC POWER, HE SAW A SMALL PUFF OF SMOKE COMING FROM THE TRANSFORMER AREA. HE FOUND 120 AC INTO THE TRANSFORMER AND 18 VAC OUT FROM THE TRANSFORMER, BUT 0 VDC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |