HERCULES 3 STAGE BALLOON ESOPHAGEAL
Report
- Report Number
- 1037905-2011-00339
- Event Type
- Malfunction
- Date Received
- May 27, 2011
- Date of Event
- April 28, 2011
- Report Date
- April 29, 2011
- Manufacturer
- WILSON-COOK MEDICAL INC
- Product Code
- KNQ
- PMA / PMN Number
- K061937
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION: DURING A VISUAL EXAMINATION OF THE BALLOON MATERIAL, WE CONFIRMED THE PRESENCE OF A LARGE SPLIT IN THE BALLOON MATERIAL. THE SPLIT IS POSITIONED LENGTH-WISE ON THE BALLOON AND IS APPROX THE SAME LENGTH OF THE BALLOON. THE CATHETER TUBING DOES NOT EXHIBIT ANY STRETCHED OR DAMAGED AREAS. DURING A FUNCTIONAL EVALUATION OF THE PRODUCT SAID TO BE INVOLVED, A COOK QUANTUM INFLATION DEVICE FILLED WITH WATER WAS USED IN AN ATTEMPT TO INFLATE THE BALLOON. WHEN THE BALLOON IS INFLATED WITH WATER, WATER EXITS THE BALLOON MATERIAL AT THE LOCATION OF THE LARGE SPLIT. THE BALLOON WILL NOT HOLD PRESSURE AND CANNOT PERFORM DILATION IN THIS CONDITION. A MANUFACTURING DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORT WAS NOT OBSERVED DURING OUR LABORATORY EVALUATION. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. A SAMPLE TEST FROM THIS LOT COULD NOT BE CONDUCTED BECAUSE NONE OF THE PRODUCT FROM THIS LOT REMAINED IN INVENTORY. CONCLUSION: THE ADDITIONAL INFO PROVIDED INDICATED THE BALLOON DILATOR DID NOT RECEIVE NEGATIVE PRESSURE PRIOR TO ADVANCEMENT THROUGH THE ENDOSCOPE. THIS IS THE MOST LIKELY CAUSE FOR THE REPORTED OBSERVATION. THE INSTRUCTIONS FOR USE DIRECT THE USER TO APPLY NEGATIVE PRESSURE TO THE CATHETER TO FACILITATE PASSAGE THROUGH THE ENDOSCOPE. THE APPLICATION OF NEGATIVE PRESSURE WILL ASPIRATE ALL RESIDUAL AIR FROM THE BALLOON AND EASE ENDOSCOPIC ADVANCEMENT. NEGATIVE PRESSURE WILL ALSO AID IN BALLOON PRESERVATION AND OPTIMIZE BALLOON PERFORMANCE. A POSSIBLE CONTRIBUTING FACTOR TO BALLOON DAMAGE IS INADEQUATE LUBRICATION OF THE BALLOON WITH A LUBRICATING AGENT. THE INSTRUCTIONS FOR USE DIRECT THE USER TO APPLY A LUBRICATING AGENT TO THE BALLOON TO FACILITATE PASSAGE THROUGH THE ENDOSCOPE ACCESSORY CHANNEL. THIS ACTIVITY WILL AID IN ENDOSCOPIC ADVANCEMENT AND BALLOON PRESERVATION. DAMAGE TO THE BALLOON MATERIAL CAN OCCUR IF THE BALLOON IS INFLATED PRIOR TO ADVANCEMENT THROUGH THE ENDOSCOPE OR IF THE BALLOON IS INFLATED WHILE PARTIALLY OR FULLY INSIDE THE ACCESSORY CHANNEL OF THE ENDOSCOPE. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING PRECAUTION: "DO NOT PRE-INFLATE THE BALLOON." THE INSTRUCTIONS FOR USE DIRECT THE USER TO ENSURE THE BALLOON IS COMPLETELY VISUALIZED AND POSITIONED BEFORE INFLATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING WARNING: "DURING DILATION, DO NOT INFLATE BALLOON BEYOND THE MAXIMUM INDICATED INFLATION PRESSURE." OVER INFLATION CAN CAUSE DAMAGE TO THE BALLOON DILATOR. ANOTHER POSSIBLE CONTRIBUTING FACTOR IS USING A COMPROMISED INFLATION DEVICE TO INFLATE THE BALLOON. IF THE PRESSURE READING OF THE INFLATION DEVICE IS INACCURATE, THIS COULD CONTRIBUTE TO OVER INFLATION, POSSIBLY RESULTING IN DAMAGE TO THE BALLOON MATERIAL. PRIOR TO DISTRIBUTION, ALL HERCULES 3 STAGE ESOPHAGEAL BALLOON DILATORS ARE SUBJECTED TO A VISUAL EXAMINATION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
IN PREPARATION FOR A PROCEDURE, THE USER SELECTED A COOK HERCULES 3 STAGE ESOPHAGEAL BALLOON. WHILE TRYING TO INFLATE THE BALLOON IN THE ESOPHAGUS, THE BALLOON RUPTURED. A FOREIGN OBJECT DID NOT HAVE TO BE RETRIEVED FROM THE PT. NO ADDITIONAL MEDICAL PROCEDURES WERE REQUIRED DUE TO THIS OCCURRENCE. THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OBSERVATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HERCULES 3 STAGE BALLOON ESOPHAGEAL | KNQ, ESOPHAGEAL DILATOR | KNQ | WILSON-COOK MEDICAL INC | W2982616 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DILATION SYRING (UNKNOWN MAKE AND MODEL)| OLYMPUS 180 ENDOSCOPE (UNKNOWN MODEL)| BOSTON SCIENTIFIC ALLIANCE INFLATION GUN |