UNICEL DXC 800 SYNCHRON SYSTEM
Report
- Report Number
- 2050012-2011-02566
- Event Type
- Malfunction
- Date Received
- June 24, 2011
- Date of Event
- April 15, 2010
- Report Date
- April 21, 2010
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- CGA
- PMA / PMN Number
- K042291
- Removal / Correction Number
- Z-2388-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
NO SERVICE CALL WAS INITIATED OR SYSTEM EVALUATION CONDUCTED BY THE MANUFACTURER. THE CUSTOMER CONTACT REPRESENTATIVE DISCUSSED SEVERAL POSSIBLE CAUSES OF THE EVENT WITH THE CUSTOMER INCLUDING PROPER OUT-GASSING OF THE REAGENT AND PROPER MODULE COVER SEATING. THE CUSTOMER DOES NOT PROPERLY SECURE THE SYSTEM COVERS; ACCORDINGLY, THE CUSTOMER CONTACT REPRESENTATIVE SUGGESTED THAT THE CUSTOMER PROPERLY SECURE THE COVERS. SINCE NO EXAMINATION OF THE SYSTEM WAS CONDUCTED A DEFINITIVE ROOT CAUSE OF THIS EVENT COULD NOT BE DETERMINED. THIS IS ONE OF THREE MEDWATCH REPORTS AS THE CUSTOMER REPORTED THREE SEPARATE INCIDENTS. REFERENCE THE MDR NUMBERS BELOW FOR ALL ASSOCIATED EVENTS: MDR # 2050012-2010-00252, MDR # 2050012-2011-02565. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS CONDUCTED BETWEEN (B)(4) 2008 AND (B)(4) 2010 FOR ADDITIONAL REPORTABLE EVENTS.
CUSTOMER REPORTED THAT AN ERRONEOUSLY LOW GLUCOSE RESULT WAS GENERATED BY THE UNICEL DXC 800 SYNCHRON SYSTEM. THE SYSTEM GENERATED CRITICAL RE-RUN FEATURE WAS TRIGGERED AND RETURNED A RESULT WITHIN EXPECTATIONS. THE SAMPLE WAS RE-TESTED AND RETURNED A RESULT WITHIN EXPECTATIONS AND REPORTED OUT OF THE LABORATORY. QUALITY CONTROLS PRIOR TO AND AFTER THE EVENT WERE WITHIN LABORATORY SPECIFICATIONS. THE ORIGINAL RESULT WAS NOT REPORTED OUTSIDE THE LABORATORY; ACCORDINGLY, THERE WAS NO CHANGE TO THE PATIENTS' CARE OR TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 800 SYNCHRON SYSTEM | CGA | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | LX GLUCOSE 2X2L |