FDA Adverse Event
Malfunction
Summary report: N
7900
MDR report key: 2172182
·
Received June 15, 2011
Report
- Report Number
- 9680959-2011-01442
- Event Type
- Malfunction
- Date Received
- June 15, 2011
- Date of Event
- June 8, 2011
- Report Date
- June 15, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE FUSE IN VIDEO AC POWER DISTRIBUTION CHASSIS WAS REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD TAKE THE FLUOROSCOPIC X-RAY BUT WOULD NOT DISPLAY THE IMAGE. NO PT INJURY WAS REPORTED.