FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND HIGH PROFILE

MDR report key: 21721369 · Received March 28, 2025

Report

Report Number
1645337-2025-03327
Event Type
Injury
Date Received
March 28, 2025
Date of Event
December 27, 2024
Manufacturer
MENTOR TEXAS
Product Code
FWM
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. D4: UDI: AS THE CATALOG/MODEL NUMBER WAS NOT PROVIDED, THE (01) GTIN IS NOT AVAILABLE. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON APR 8, 2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE IMPACTED PRODUCT WAS IDENTIFIED AS A SALINE MENTOR SMOOTH ROUND HIGH PROFILE 310CC BREAST IMPLANT, LOT NUMBER 5780290. DEVICE EVALUATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION, MICROSCOPIC EXAMINATION, AND THICKNESS MEASUREMENT OF THE RETURNED DEVICE. DURING VISUAL ANALYSIS OF THE RETURNED DEVICE SM HPS DV 310CC A TEAR NEAR THE VALVE SYSTEM WAS FOUND. MICROSCOPIC EXAMINATION WAS PERFORMED, AND THE CAUSE OF THE TEAR COULD NOT BE IDENTIFIED. THEREFORE A THICKNESS MEASUREMENT WAS CONDUCTED ON THE SHELL AT THE TEAR, AND THE THICKNESS WAS WITHIN MANUFACTURING SPECIFICATIONS. MENTOR ALSO PERFORMED A MANUFACTURING RECORD EVALUATION RELATED TO THE REPORTED COMPLAINT FOR THE FINISHED DEVICE LOT NUMBER. NO MANUFACTURING NON-CONFORMANCES WERE IDENTIFIED AS PART OF THIS EVALUATION. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: ACCORDING WITH THE PRODUCT INSERT DATA SHEET, THE VALVE SYSTEM CAN BE DAMAGED BY IMPROPER USE OF THE FILL-TUBE STYLET. CARE SHOULD BE TAKEN THAT THE STYLET ENTERS THE VALVE SMOOTHLY. USE THE THUMB AND FOREFINGER TO STABILIZE/SUPPORT THE VALVE SEAT AND GENTLY PUSH THE STYLET TIP INTO THE VALVE OPENING. OVERSTRESSING THE VALVE MATERIAL MAY RESULT IN PUNCTURES OR TEARS AND SUBSEQUENT DEFLATION MAY OCCUR. USE ONLY THE FILL TUBE STYLET PROVIDED WITH THIS PRODUCT. TAKE CARE NOT TO PUNCTURE THE DIAPHRAGM VALVE OR THE SHELL WITH THE STYLET TIP. CARE MUST ALSO BE TAKEN WHEN THE FILL TUBE STYLET IS REMOVED TO PREVENT DAMAGE TO THE VALVE ASSEMBLY. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. D4: UDI: AS THE SERIAL NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION PRIMARY WITH UNSPECIFIED MENTOR SALINE BREAST IMPLANTS AND EXPERIENCED DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT REMOVAL AND REPLACEMENT WITH SIMILAR DEVICES ON (B)(6) 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
712477 MENTOR SMOOTH ROUND HIGH PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5780290

Patients

Seq Age Sex Outcome Treatment
1 45 YR Female Required Intervention