SWAN-GANZ CCOMBO CCO/SVO2/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHIELD
Report
- Report Number
- 2015691-2011-15908
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 20, 2011
- Report Date
- June 20, 2011
- Manufacturer
- EDWARDS LIFESCIENCES, PR
- Product Code
- DYG
- PMA / PMN Number
- K934742
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE CATHETER WAS RECEIVED WITH THE CONTAMINATION SHIELD ATTACHED TO THE CATHETER, THE PROXIMAL END WAS APPROXIMATELY 89CM FROM THE TIP. AN EDWARDS PRESSURE TRANSDUCER AND ASSOCIATED TUBING WERE ATTACHED TO PA DISTAL HUB. TESTING WAS DONE WITH THE RETURNED DPT TUBING CONNECTED TO THE HUB. THE CAL-CUP WAS NOT RETURNED, TESTING WAS DONE USING A LAB CAL-CUP. UPON REMOVAL OF THE CONTAMINATION SHIELD, THE CATHETER BODY APPEARED COLLAPSED APPROXIMATELY 89CM FROM THE TIP. THE CATHETER PASSED IN-VITRO CALIBRATION ON BOTH VIGILANCE I AND II MONITORS. THE CATHETER ALSO PASSED TRANSMISSION AND CROSS-TALK TESTING. THE CATHETER RAN CCO ON VIGILANCE I AND VIGILANCE II MONITORS FOR 5 MINUTES WITH NO ERROR MESSAGES. NO VISIBLE INCONSISTENCIES WERE OBSERVED ON EEPROM DATA. THE THERMISTOR AND THERMAL FILAMENT CIRCUIT WERE CONTINUOUS. THE THERMISTOR READ 37.1 C WHEN SUBMERGED IN A 37.0 C WATER BATH. THERE WERE NO OPEN OR INTERMITTENT CONDITIONS. BOTH THERMISTOR AND THERMAL FILAMENT CONNECTORS WERE OPENED WITH NO VISIBLE INCONSISTENCIES. THE BALLOON INFLATED CLEARLY AND CONCENTRICALLY AND REMAINED INFLATED FOR 5 MINUTES WITHOUT LEAKAGE. A 10CC SYRINGE WAS CONNECTED TO EACH OF THE LUMENS BEING TESTED. EACH ASSOCIATED PORT WAS COVERED WITH A FINGERTIP, ENTIRE CATHETER WAS IMMERSED IN WATER AND PRESSURIZED WITH AIR BY COMPRESSING THE SYRINGE PLUNGER. ALL THROUGH LUMENS WERE PATENT WITHOUT LEAKAGE OR OCCLUSION. NO VISIBLE DAMAGE WAS OBSERVED FROM THE OPTICAL MODULE CONNECTOR OR RETURNED MONOJECT 1.5CC LIMITED VOLUME SYRINGE. THE PA DISTAL HUB AND PROX INJECTATE HUB APPEARED TO HAVE BEEN DAMAGED FROM CLAMPS. CARDIAC OUTPUT TESTING WAS DONE ON VIGILANCE I AND II MONITORS. DURING TESTING OF THE RETURNED PRESSURE TRANSDUCER, THE DPT ZEROED AND SENSED PRESSURE; NO FAILURES WERE NOTED. THE COMPLAINT COULD NOT BE CONFIRMED DURING TESTING; NO PRODUCT PROBLEMS WERE IDENTIFIED. IT IS NOT KNOWN IF SOME PROCEDURAL FACTORS MAY HAVE CONTRIBUTED TO THIS EVENT. NO ACTIONS WILL BE TAKEN AT THIS TIME. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED.
IT WAS REPORTED THAT CCO AND SVO2 MEASUREMENTS WERE "BAD" IMMEDIATELY AFTER INSERTION BY THE CARDIOLOGIST. ENGINEERING DID NOT PROVIDE DIRECT TROUBLESHOOTING SUPPORT. FOLLOW UP INDICATED THAT WRONG SVO2 AND CCO NUMBERS WERE "RIDICULOUS", THE NUMBERS LESS THAN 4 WERE OBSERVED ON THE MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SWAN-GANZ CCOMBO CCO/SVO2/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHIELD | CONTINOUS CARDIAC OUPTUT CATHETER | DYG | EDWARDS LIFESCIENCES, PR | 746HF8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |