FDA Adverse Event Malfunction Summary report: N

SWAN-GANZ CCOMBO CCO/SVO2/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHIELD

MDR report key: 2172079 · Received July 22, 2011

Report

Report Number
2015691-2011-15908
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 20, 2011
Report Date
June 20, 2011
Manufacturer
EDWARDS LIFESCIENCES, PR
Product Code
DYG
PMA / PMN Number
K934742
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE CATHETER WAS RECEIVED WITH THE CONTAMINATION SHIELD ATTACHED TO THE CATHETER, THE PROXIMAL END WAS APPROXIMATELY 89CM FROM THE TIP. AN EDWARDS PRESSURE TRANSDUCER AND ASSOCIATED TUBING WERE ATTACHED TO PA DISTAL HUB. TESTING WAS DONE WITH THE RETURNED DPT TUBING CONNECTED TO THE HUB. THE CAL-CUP WAS NOT RETURNED, TESTING WAS DONE USING A LAB CAL-CUP. UPON REMOVAL OF THE CONTAMINATION SHIELD, THE CATHETER BODY APPEARED COLLAPSED APPROXIMATELY 89CM FROM THE TIP. THE CATHETER PASSED IN-VITRO CALIBRATION ON BOTH VIGILANCE I AND II MONITORS. THE CATHETER ALSO PASSED TRANSMISSION AND CROSS-TALK TESTING. THE CATHETER RAN CCO ON VIGILANCE I AND VIGILANCE II MONITORS FOR 5 MINUTES WITH NO ERROR MESSAGES. NO VISIBLE INCONSISTENCIES WERE OBSERVED ON EEPROM DATA. THE THERMISTOR AND THERMAL FILAMENT CIRCUIT WERE CONTINUOUS. THE THERMISTOR READ 37.1 C WHEN SUBMERGED IN A 37.0 C WATER BATH. THERE WERE NO OPEN OR INTERMITTENT CONDITIONS. BOTH THERMISTOR AND THERMAL FILAMENT CONNECTORS WERE OPENED WITH NO VISIBLE INCONSISTENCIES. THE BALLOON INFLATED CLEARLY AND CONCENTRICALLY AND REMAINED INFLATED FOR 5 MINUTES WITHOUT LEAKAGE. A 10CC SYRINGE WAS CONNECTED TO EACH OF THE LUMENS BEING TESTED. EACH ASSOCIATED PORT WAS COVERED WITH A FINGERTIP, ENTIRE CATHETER WAS IMMERSED IN WATER AND PRESSURIZED WITH AIR BY COMPRESSING THE SYRINGE PLUNGER. ALL THROUGH LUMENS WERE PATENT WITHOUT LEAKAGE OR OCCLUSION. NO VISIBLE DAMAGE WAS OBSERVED FROM THE OPTICAL MODULE CONNECTOR OR RETURNED MONOJECT 1.5CC LIMITED VOLUME SYRINGE. THE PA DISTAL HUB AND PROX INJECTATE HUB APPEARED TO HAVE BEEN DAMAGED FROM CLAMPS. CARDIAC OUTPUT TESTING WAS DONE ON VIGILANCE I AND II MONITORS. DURING TESTING OF THE RETURNED PRESSURE TRANSDUCER, THE DPT ZEROED AND SENSED PRESSURE; NO FAILURES WERE NOTED. THE COMPLAINT COULD NOT BE CONFIRMED DURING TESTING; NO PRODUCT PROBLEMS WERE IDENTIFIED. IT IS NOT KNOWN IF SOME PROCEDURAL FACTORS MAY HAVE CONTRIBUTED TO THIS EVENT. NO ACTIONS WILL BE TAKEN AT THIS TIME. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT CCO AND SVO2 MEASUREMENTS WERE "BAD" IMMEDIATELY AFTER INSERTION BY THE CARDIOLOGIST. ENGINEERING DID NOT PROVIDE DIRECT TROUBLESHOOTING SUPPORT. FOLLOW UP INDICATED THAT WRONG SVO2 AND CCO NUMBERS WERE "RIDICULOUS", THE NUMBERS LESS THAN 4 WERE OBSERVED ON THE MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SWAN-GANZ CCOMBO CCO/SVO2/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHIELD CONTINOUS CARDIAC OUPTUT CATHETER DYG EDWARDS LIFESCIENCES, PR 746HF8

Patients

Seq Age Sex Outcome Treatment
1