FDA Adverse Event Malfunction Summary report: N

SWAN-GANZ CCOMBO CCO/SVO2/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHIELD

MDR report key: 2172033 · Received July 22, 2011

Report

Report Number
2015691-2011-15907
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 20, 2011
Report Date
June 20, 2011
Manufacturer
EDWARDS LIFESCIENCES, PR
Product Code
DYG
PMA / PMN Number
K934742
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE CATHETER WAS RECEIVED WITH THE CONTAMINATION SHIELD ATTACHED TO THE CATHETER AT THE PROXIMAL END APPROXIMATELY 94CM FROM THE TIP. THE CAL-CUP WAS NOT RETURNED, TESTING WAS DONE USING A LAB CAL-CUP. THE CATHETER PASSED IN-VITRO CALIBRATION ON BOTH VIGILANCE I AND II MONITORS. THE CATHETER ALSO PASSED TRANSMISSION AND CROSS-TALK TESTING. THE CATHETER RAN CCO ON VIGILANCE I AND VIGILANCE II MONITORS FOR 5 MINUTES WITH NO ERROR MESSAGES. NO VISIBLE INCONSISTENCIES WERE OBSERVED ON EEPROM DATA. THE THERMISTOR AND THERMAL FILAMENT CIRCUIT WERE CONTINUOUS. THE READ 37.1 C WHEN SUBMERGED IN A 37.0 C WATER BATH. THERE WERE NO OPEN OR INTERMITTENT CONDITIONS. BOTH THERMISTOR AND THERMAL FILAMENT CONNECTORS WERE OPENED WITH NO VISIBLE INCONSISTENCIES. THE SYRINGE WAS NOT RETURNED, TESTING WAS DONE USING LAB SYRINGE. THE BALLOON INFLATED CLEAR AND CONCENTRIC AND REMAINED INFLATED FOR 5 MINUTES. WITHOUT LEAKAGE. A 10CC SYRINGE WAS CONNECTED TO EACH OF THE LUMENS BEING TESTED. EACH ASSOCIATED PORT WAS COVERED WITH A FINGERTIP; ENTIRE CATHETER WAS IMMERSED IN WATER AND PRESSURIZED WITH AIR BY COMPRESSING THE SYRINGE PLUNGER. ALL THROUGH LUMENS WERE PATENT WITHOUT LEAKAGE OR OCCLUSION. NO VISIBLE DAMAGE WAS OBSERVED ON THE OPTICAL MODULE CONNECTOR OR CATHETER BODY. VISUAL EXAMINATION WAS PERFORMED UNDER 10 X MAGNIFICATIONS. THE COMPLAINT COULD NOT BE CONFIRMED DURING TESTING; NO PRODUCT PROBLEMS WERE IDENTIFIED. IT IS NOT KNOWN IF SOME PROCEDURAL FACTORS MAY HAVE CONTRIBUTED TO THIS EVENT. NO ACTIONS WILL BE TAKEN AT THIS TIME. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CARDIAC INDEX (CI) AND CARDIAC OUTPUT (CO) WOULD DROP TO ZERO. IT WAS FURTHER INDICATED THAT THE INCORRECT SVO2 MEASUREMENTS ALSO PROMPTED NUMEROUS RECALIBRATIONS BY THE NURSE. ALL CABLES AND CCO BOXES WERE EXCHANGED WITH KNOWN GOOD EQUIPMENT. ENGINEERING DID NOT PROVIDE DIRECT TROUBLESHOOTING SUPPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SWAN-GANZ CCOMBO CCO/SVO2/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHIELD CONTINOUS CARDIAC OUPTUT CATHETER DYG EDWARDS LIFESCIENCES, PR 746HF8

Patients

Seq Age Sex Outcome Treatment
1