FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2172027 · Received July 22, 2011

Report

Report Number
6000001-2011-13448
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 24, 2011
Report Date
June 24, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.48.92. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH INOPERATIVE "SELECT" BUTTON ON THE CHANNEL A PUMP HEAD MODULE KEYPAD WAS CONFIRMED DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN. THIS CONDITION WAS CAUSED BY FLUID INGRESS RESIDUE AND DAMAGE TO PHM KEYPAD FLEX CABLE. THE PUMP HEAD MODULE KEYPAD WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED TO BAXTER (B)(4), A COLLEAGUE INFUSION PUMP WITH AN INOPERATIVE "SELECT" BUTTON ON THE CHANNEL A PUMP HEAD MODULE KEYPAD. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1