ELEVATOR HEIDBRINK DE 13/14
Report
- Report Number
- 8010704-2006-00001
- Event Type
- Malfunction
- Date Received
- April 13, 2006
- Date of Event
- February 28, 2006
- Report Date
- February 28, 2006
- Manufacturer
- NOPA INSTRUMENTS, MEDIZINTECHNIK GMBH
- Product Code
- EMJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- DENTIST
Narratives
THE INSTRUMENT WAS NOT RETURNED TO THE MFR. WHEN QUESTIONED ON THE TELEPHONE ON (B)(6) 2006, THE INITIAL RPTR STATED THAT HE DOES NOT KNOW WHETHER THE INSTRUMENT WAS DISCARDED OR NOT, SINCE HE WORKS AT A MEDICAL CENTER AND IS NOT IN CHARGE OF THESE MATTERS. HE WILL HAVE THIS CHECKED AND HAS PROMISED TO GET BACK TO (B)(4) AS SOON AS MORE INFO IS AVAILABLE. HE ALSO HAS NO IDEA FROM WHICH DISTRIBUTOR OR MFR - OF THE HUNDREDS WHO MARKET THIS TYPE OF PRODUCT-- THE INSTRUMENT WAS PURCHASED NOR WHEN. (B)(4) IS ONE OF MANY DISTRIBUTORS HIS MEDICAL CENTER WORKS WITH AND WAS CHOSEN BY HIM COMPLETELY AT RANDOM SIMPLY TO INQUIRE ABOUT THE PT BEING ABLE TO HAVE AN MRI OR NOT. (B)(4) IN TURN USED TO ACQUIRE THIS TYPE OF INSTRUMENT FROM ANOTHER MFR AND/OR REPACKAGER/RELABELER BEFORE (B)(4). SINCE THE INSTRUMENT HAS NOT BEEN RETURNED, IT WAS IMPOSSIBLE TO DETERMINE THE DATE OF PURCHASE. (B)(4), AS THE CURRENT SUPPLIER, WAS NOTIFIED. ALTHOUGH NO CERTAINTY WHATSOEVER EXISTS THAT THE INSTRUMENT INVOLVED IN THE EVENT ORIGINALLY CAME FROM (B)(4), AFTER RECEIVING INFO ABOUT THE EVENT, (B)(4) PROCEEDED. ALL INSTRUMENTS OF THE TYPE INVOLVED WERE SUBJECTED TO FATIGUE, ROCKWELL AND BREAKAGE TESTS. ON (B)(6) 2006, ALL TESTS WERE COMPLETED WITHIN SPECIFICATIONS.
THE ELEVATOR TIP BROKE OFF IN THE PT'S BONE DURING A TOOTH EXTRACTION. AN X-RAY SHOWED THAT A SMALL PIECE OF THE BROKEN TIP IS LODGED IN THE PT'S BONE. WHEN QUESTIONED ABOUT THIS ON THE TELEPHONE, THE DOCTOR STATED THAT HE DECIDED NOT TO REMOVE THE BROKEN PIECE SINCE IT IS LOCATED TOO CLOSE TO A NERVE AND HE DOES NOT "WISH TO GO IN THERE." THE DOCTOR'S ONLY CONCERN IS REGARDING THE ADVISABILITY OF SUBJECTING THE PT TO AN MRI. NO DAMAGE OR HARM HAS COME TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ELEVATOR HEIDBRINK DE 13/14 | DENTAL ROOT TIP ELEVATOR | EMJ | NOPA INSTRUMENTS, MEDIZINTECHNIK GMBH | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |