PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2011-05080
- Event Type
- Injury
- Date Received
- July 22, 2011
- Date of Event
- June 28, 2011
- Report Date
- June 28, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE DEVICE WAS REPORTED TO HAVE BEEN DISCARDED. WITHOUT THE DEVICE EVALUATION, A CAUSE FOR THE REPORTED CUFF MISS COULD NOT BE DETERMINED. A CUFF MISS CAN BE AFFECTED BY NUMEROUS FACTORS INCLUDING, BUT NOT LIMITED TO, MANUFACTURING ISSUES, USER TECHNIQUE, OR PATIENT ANATOMY. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, THE NEEDLE TRAJECTORY OF EVERY DEVICE IS CHECKED DURING MANUFACTURING. IN ADDITION, A SAMPLING OF FINISHED DEVICES IS DESTRUCTIVELY TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE. FAILURE TO POSITION AND MAINTAIN THE DEVICE AT A 45 DEGREE ANGLE THROUGHOUT DEPLOYMENT, ROTATING THE DEVICE AT ANY POINT DURING PLUNGER/NEEDLE DEPLOYMENT, AGGRESSIVELY DEPLOYING OR REMOVING THE PLUNGER AND / OR INADEQUATE POSITIONING OF THE FOOT AGAINST THE ARTERIAL WALL CAN BE CONTRIBUTING FACTORS. HOWEVER, INFORMATION IN REGARDS TO THE USER TECHNIQUE WAS NOT PROVIDED. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS THAT WOULD HAVE AFFECTED THE REPORTED NEEDLE TO CUFF MISS EVENT. A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS PERFORMED AND NO PRODUCT QUALITY DEFICIENCY WAS IDENTIFIED.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE FIRST PERCLOSE PROGLIDE DEVICE IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PERCLOSE PROGLIDE DEVICE AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. ANOTHER PROGLIDE WAS USED WITH THE SAME RESULTS. HEMOSTASIS WAS ACHIEVED USING A THIRD PROGLIDE DEVICE. THERE WAS NO ADVERSE PATIENT SEQUELA. REPORTEDLY, THE PHYSICIAN IS TRAINED IN THE USE OF THE DEVICE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 030336H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SHEATH: 6F |