ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-05112
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Report Date
- May 16, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: DURING EVALUATION THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION AND MOISTURE DAMAGE WAS FOUND ON THE FORCE SENSOR ASSEMBLY. DURING TESTING THE LOAD CARTRIDGE STEP DID NOT DETECT THE CARTRIDGE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6). RECALL #: 2531779-03/24/2010-003-R.
A FAMILY MEMBER REPORTED THAT THE PUMP DISPENSED INSULIN DURING THE LOAD CARTRIDGE PHASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR |