FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2171709 · Received July 22, 2011

Report

Report Number
6000001-2011-13339
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 22, 2011
Report Date
June 23, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON FURTHER INVESTIGATION, IT WAS DETERMINED THAT USER ERROR HAS CONTRIBUTED TO THIS EVENT.

Additional Manufacturer Narrative · 1

THE CONDITION OF DAMAGED BATTERY IS CONFIRMED DUE TO DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4)

Description of Event or Problem · 1

THIS IS A REPORT OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY, WHICH COULD HAVE CAUSED AN INTERRUPTION OF DELIVERY. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90 CATEGORIZED AS REMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1