COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-13337
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A BROKEN TUBE RELEASE WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THE REPORTED PROBLEM WAS DETERMINED TO BE A BROKEN MANUAL TUBE RELEASE (MTR) KNOB. APPROPRIATE ACTION WAS TAKEN TO CORRECT THE REPORTED PROBLEM BY REPLACING THE MTR KNOB. ADDITIONAL INFORMATION: THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.
THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. UPON COMPLETION OF BAXTER'S INVESTIGATION A FOLLOW UP MED WATCH WILL BE SUBMITTED. (B)(4).
THIS IS A REPORT OF A COLLEAGUE INFUSION PUMP WITH A BROKEN TUBE RELEASE, WHICH COULD HAVE CAUSED AN INTERRUPTION OF DELIVERY. THE REPORTED CONDITION OCCURRED DURING PROGRAMMING/SETUP. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. THE SOFTWARE VERSION IS CURRENTLY NOT KNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |