FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2171637 · Received July 21, 2011

Report

Report Number
2531779-2011-05100
Event Type
Malfunction
Date Received
July 21, 2011
Report Date
June 21, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE KEYPAD BUTTONS HAVE BEEN STICKING. HE STATED THAT THE BUTTONS SOMETIMES REQUIRE THREE TO FOUR PRESSES TO OBTAIN A RESPONSE, AND OTHER TIMES ARE COMPLETELY UNRESPONSIVE. HE CONFIRMED THAT THE RUBBER KEYPAD IS INTACT AND STATED THAT THE BUTTONS STILL CLICK AND SPRING BACK AS EXPECTED. THE PATIENT REPORTED THAT HE WEARS THE PUMP IN HIS POCKET, AND CLEANS THE PUMP WITH AN ALCOHOL SWAB.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 47 YR