FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 2171621 · Received July 21, 2011

Report

Report Number
2124215-2011-09887
Event Type
Injury
Date Received
July 21, 2011
Date of Event
May 1, 2008
Report Date
June 22, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED AND IS CURRENTLY BEING ANALYZED. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE IMPLANT PROCEDURE THIS IMPLANTABLE LEFT VENTRICULAR (LV) LEAD CAUSED A PERFORATION IN A LATERAL VEIN. THE LEAD WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4543

Patients

Seq Age Sex Outcome Treatment
1 76 YR Life Threatening 0184| 4538| H217| 4543| 4135| 4554