FDA Adverse Event Injury Summary report: N

TELIGEN

MDR report key: 2171613 · Received July 21, 2011

Report

Report Number
2124215-2011-10232
Event Type
Injury
Date Received
July 21, 2011
Date of Event
June 3, 2011
Report Date
June 18, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
PMA / PMN Number
P010012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

ALL AVAILABLE INFORMATION INDICATES THAT THE PRODUCTS REMAIN IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED ADDITIONAL INFORMATION THREE MONTHS LATER THAT THIS RIGHT VENTRICULAR DEFIBRILLATION LEAD AND IMPLANTABLE CARDIOVERTER DEFIBRILLATOR CONTINUE TO DISPLAY HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS. THE PRODUCTS REMAIN IMPLANTED AND WILL CONTINUE TO BE MONITORED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR DEFIBRILLATION LEAD AND IMPLANTABLE CARDIOVERTER DEFIBRILLATOR DISPLAYED HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS AFTER LESS THAN TWO MONTHS OF IMPLANT TIME. UPON DEVICE INTERROGATION, THE LEAD IMPEDANCE MEASUREMENTS WERE NORMAL. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THIS PATIENT UNDERWENT A REVISION PROCEDURE. IT WAS REVEALED THAT THE RV LEAD WAS NOT FULLY INSERTED IN THE HEADER OF THE ICD. THE LEAD WAS RE-INSERTED SUCCESSFULLY AND NO FURTHER COMPLICATIONS WERE OBSERVED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS PRODUCT REMAINS IN-SERVICE.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THIS SYSTEM CONTINUED TO EXHIBIT OUT OF RANGE IMPEDANCE MEASUREMENTS. THE PATIENT WAS BROUGHT INTO THE CLINIC FOR EVALUATION AND THERE WAS ONE EPISODE OF NOISE STORED. AN X-RAY WAS PERFORMED AND IT WAS REPORTED THAT THERE WAS A POTENTIAL INCOMPLETE CONNECTION ISSUE. THE PATIENT WAS TO UNDERGO A REVISION PROCEDURE IN THE NEAR FUTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND E102

Patients

Seq Age Sex Outcome Treatment
1 41 YR Hospitalization| L| R E102| 0181