FLEXTEND II
Report
- Report Number
- 2124215-2011-09763
- Event Type
- Injury
- Date Received
- July 21, 2011
- Date of Event
- May 31, 2011
- Report Date
- July 28, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS INVESTIGATION WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD AND LEFT VENTRICULAR LEAD WERE EXPLANTED DUE TO A DISLODGEMENT. IT WAS ALSO NOTED THAT THE LEFT VENTRICULAR LEAD SHOWED A CHANGE IN PACING IMPEDANCES. THE LEADS WILL BE RETURNED. NO ADVERSE PATIENT EFFECTS WERE REPORTED FOLLOWING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXTEND II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4096 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |