FDA Adverse Event Injury Summary report: N

FLEXTEND

MDR report key: 2171566 · Received July 21, 2011

Report

Report Number
2124215-2011-11417
Event Type
Injury
Date Received
July 21, 2011
Date of Event
June 3, 2011
Report Date
June 3, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE LEAD WAS SURGICALLY ABANDONED AND WAS NOT RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEAD WAS EXHIBITING IMPEDANCE MEASUREMENTS LESS THAN 200 OHMS. ADDITIONALLY, OVERSENSING WAS NOTED. A LEAD REVISION WAS PERFORMED AND THE LEAD WAS REMOVED FROM SERVICE AND SUCCESSFULLY REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXTEND IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4088

Patients

Seq Age Sex Outcome Treatment
1 84 YR 4137| 4087| 1298| S603| 4088