FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 2171542 · Received July 21, 2011

Report

Report Number
2124215-2011-10375
Event Type
Injury
Date Received
July 21, 2011
Date of Event
April 12, 2011
Report Date
July 6, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. VISUAL INSPECTION OF THE LEAD REVEALED THE COILS WERE COMPRESSED 19 CENTIMETERS (CM) FROM THE TERMINAL PIN. A CUT IN THE INSULATION WAS ALSO NOTED 39.5 CM AND 57 CM FROM THE TERMINAL PIN. THE DAMAGE OBSERVED LIKELY OCCURRED DURING THE EXPLANT PROCEDURE. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS OF THE CONDUCTIVE PATHS WERE WITHIN NORMAL LIMITS. PRESSURE TESTS WERE NOT SUCCESSFUL DUE TO THE CUTS IN THE LEAD INSULATION. MICROSCOPIC INSPECTIONS OF THE ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS.

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS REPORT WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR PACING LEAD EXHIBITED LOSS OF CAPTURE. A DECREASE IN PACING IMPEDANCE MEASUREMENTS AND AMPLITUDE MEASUREMENTS WAS ALSO OBSERVED. A LEAD REVISION WAS PERFORMED AND FLUOROSCOPY SHOWED THE LEAD HAD DISLODGED. THE LEAD WAS EXPLANTED, SUCCESSFULLY REPLACED, AND RETURNED FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4518

Patients

Seq Age Sex Outcome Treatment
1 88 YR Required Intervention 4518| 4135| N118| 0184