ACCOLADE MRI EL DR
Report
- Report Number
- 2124215-2025-19230
- Event Type
- Malfunction
- Date Received
- March 27, 2025
- Date of Event
- March 14, 2025
- Report Date
- October 20, 2025
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- LWP
- UDI-DI
- 00802526614736
- PMA / PMN Number
- P150012/S000
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- 003
Narratives
REPORT NUMBER: D1 FIELD CORRECTION FROM INGEVITY+ TO ACCOLADE MRI EL DR D2A FIELD CORRECTION FROM IMPLANTABLE LEAD TO IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) D2B FIELD CORRECTION FROM NVN TO LWP D4 FIELD CORRECTION FROM 7841, (B)(6), 7841, 02/17/2024, (B)(6), 00802526559273, (B)(4), (B)(4) TO L331, 980604, L331, 02/24/2024, 980604, 00802526559273, (B)(4), (B)(4). (B)(6). E2 FIELD CORRECTION FROM YES TO NO. E3 FIELD CORRECTION FROM OTHER HEALTH CARE PROFESSIONAL TO NON-HEALTHCARE PROFESSIONAL.
IT WAS REPORTED THAT THE ATRIAL INTRINSIC AMPLITUDE WAS OUT-OF-RANGE AND UNDERSENSING WAS OBSERVED ON THE ELECTROGRAM (EGM) FOR THIS PACEMAKER. THE PACEMAKER SYSTEM REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
IT WAS REPORTED THAT THE ATRIAL INTRINSIC AMPLITUDE WAS OUT-OF-RANGE AND UNDERSENSING WAS OBSERVED ON THE ELECTROGRAM (EGM) FOR THIS PACEMAKER. THE PACEMAKER SYSTEM REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION RECEIVED ADVISED THE STORED RIGHT VENTRICULAR AUTOMATIC THRESHOLD (RVAT) EGM SHOWED UNDERSENSING ON THE ATRIAL CHANNEL CAUSING INAPPROPRIATE VENTRICULAR PACING. CURRENTLY, THE ATRIAL SENSITIVITY IS PROGRAMMED AUTOMATIC GAIN CONTROL (AGC) AT 1.0MV (MILLIVOLTS). FURTHER REVIEW WAS RECOMMENDED. AT THIS TIME THE DEVICE REMAINS IN SERVICE AND NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1470111 | ACCOLADE MRI EL DR | IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) | LWP | BOSTON SCIENTIFIC CORPORATION | L331 | 980604 | 00802526614736 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Unknown |