FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOGUARD

MDR report key: 21713803 · Received March 27, 2025

Report

Report Number
9610048-2025-00028
Event Type
Malfunction
Date Received
March 27, 2025
Date of Event
January 5, 2025
Report Date
April 14, 2025
Manufacturer
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
Product Code
FOZ
UDI-DI
00382903818235
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 38182314 AND LOT NUMBER 4183541. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS NEITHER PHYSICAL SAMPLES NOR PICTURE SAMPLES WERE AVAILABLE FOR RETURN, A THOROUGH SAMPLE ANALYSIS COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE COULD NOT BE DETERMINED FOR THE REPORTED EVENT. HOWEVER, THE PROBABLE ROOT CAUSE COULD BE RELATED TO SPRING FORMATION DURING MANUFACTURING. THE OPERATION AND MAINTENANCE CORRECTIONS PERFORMED VARIOUS ADJUSTMENTS AND ALIGNMENTS THAT MAY HAVE BEEN RELATED TO THIS DEFECT. A CORRECTIVE AND PREVENTIVE ACTION PLAN WAS IMPLEMENTED TO PREVENT THE OCCURRENCE OF THIS TYPE OF DEFECT. HOWEVER, THE LOT REPORTED WAS MANUFACTURED PRIOR TO THIS CAPA IMPLEMENTATION. AT THIS TIME, FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY. OUR QUALITY TEAM WILL CONTINUE TO CLOSELY MONITOR THE MANUFACTURING PROCESS FOR SIGNS OF THIS POTENTIAL DEFECT AND ANY EMERGING TRENDS.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. (B)(6).

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOGUARD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER TRANSLATED FROM PORTUGUESE TO ENGLISH: ABOCATH N 22 THE SAFETY DEVICE HAS NOT RETRACTED THE NEEDLE. LOT 4183541 BRAND BD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2472081 BD INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. 4183541 00382903818235

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown