FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 2171375 · Received July 21, 2011

Report

Report Number
2124215-2011-10372
Event Type
Injury
Date Received
July 21, 2011
Date of Event
June 7, 2011
Report Date
June 7, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD EXHIBITED PACING IMPEDANCE MEASUREMENTS GREATER THAN 2000 OHMS WITH LOSS OF CAPTURE. THE PATIENT'S INTRINSIC RATE WAS 40 BEATS PER MINUTE (BPM). IT WAS DETERMINED THE LEAD HAD FRACTURED AS A RESULT OF A PATIENT FALL. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS SURGICALLY ABANDONED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4457

Patients

Seq Age Sex Outcome Treatment
1 89 YR Required Intervention S601| 4456| 1198| 4457