FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 SLIDE STAINER

MDR report key: 2171245 · Received July 1, 2011

Report

Report Number
1061932-2011-00778
Event Type
Malfunction
Date Received
July 1, 2011
Date of Event
February 25, 2008
Report Date
February 25, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
KPA
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS PER PRODUCT LABELING, BECKMAN COULTER INC URGES CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS, SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. ON 02/27/2008, THE FIELD SERVICE ENGINEER (FSE) INSPECTED AND REPLACED THE FITTING ON THE DRAIN TUBING IN BATH #2. THE FSE CHECKED THE LINE FOR BLOCKAGE. ROOT CAUSE WAS THE FITTING ON THE DRAIN TUBING IN BATH #2. REPLACING THE FITTING WAS THE CORRECTION. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW CONDUCTED BETWEEN JANUARY 1, 2008 AND OCTOBER 23, 2010 OF COMPLAINTS FOR ADDITIONAL REPORTABLE EVENTS.

Description of Event or Problem · 1

CUSTOMER REPORTED A POTENTIAL CHEMICAL HAZARD WHEN TWO LABORATORY PERSONNEL WERE SPRAYED WITH WRIGHT STAIN REAGENT FROM THE COULTER LH 750 SLIDE STAINER. THE INCIDENT OCCURRED WHEN THE BATH #2 DRAIN LINE TUBING BECAME DISCONNECTED FROM THE FITTING ON THE COULTER LH 750 SLIDE STAINER. THE TWO LABORATORY PERSONNEL WERE WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) OF LAB COATS, GLASSES AND GLOVES. THE WRIGHT STAIN REAGENT WAS SPRAYED ONTO CLOTHING AND PPE. SKIN, MUCOUS MEMBRANES OR OPEN LESIONS WERE NOT EXPOSED TO THE REAGENT. MEDICAL TREATMENT WAS NOT SOUGHT FOR EITHER LABORATORY PERSONNEL. NO REPORTS OF DEATH OR SERIOUS INJURY, AND NO AFFECT TO OPERATOR SAFETY AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER LH 750 SLIDE STAINER KPA BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK