UNICEL DXC SYNCHRON CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-02556
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Date of Event
- December 18, 2009
- Report Date
- December 18, 2009
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Removal / Correction Number
- Z-0863-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
A BCI FSE (FIELD SERVICE ENGINEER) WAS NOT DISPATCHED TO THE SITE SINCE ONE WAS REQUESTED BY THE CUSTOMER. THE BCI HOTLINE PROVIDED THE CUSTOMER WITH THE TWICE-WEEKLY FLOW CELL MAINTENANCE PROCEDURE. THE CUSTOMER CLEANED THE ISE FLOW CELL AND THIS RESOLVED THE PROBLEM. HOWEVER, A DEFINITIVE ROOT CAUSE WAS NOT DETERMINED. THIS IS ONE OF SIX SEPARATE MEDWATCH REPORTS RELATED TO THIS EVENT. ALL ASSOCIATED REPORTS ARE LISTED BELOW: 2050012-2010-00001, 2050012-2011-02551, 2050012-2011-02555, 2050012-2011-02557, 2050012-2011-02558. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS CONDUCTED FROM (B)(4), 2008 THOUGH (B)(4), 2010 FOR ADDITIONAL REPORTABLE EVENTS.
THE CUSTOMER REPORTED THE BECKMAN COULTER, INC (BCI) THAT ERRONEOUSLY LOW SODIUM (NA) AND CALCIUM (CA) RESULTS WERE OBTAINED ON THEIR UNICEL DXC 600 INSTRUMENT AND THE RESULTS WERE REPORTED OUT OF THE LABORATORY. PRIOR TO THE EVENT, THE ISE (ION-SELECTIVE ELECTRODE) SYSTEM WAS CALIBRATED AND QC (QUALITY CONTROL) RESULTS WERE FOUND TO BE OUTSIDE THE LAB'S ESTABLISHED RANGES. THE QC RUNS FOR NA AND CA WERE IMMEDIATELY REPEATED AND FOUND TO BE WITHIN THE ACCEPTABLE RANGES BUT AT THE LOW END OF THE RANGES. SAMPLES WERE ANALYZED AND SEVERAL LOW NA AND CA RESULTS WERE REPORTED OUT OF THE LAB. THE CUSTOMER RE-RAN THE PATIENT SAMPLES, FOUND THE RESULTS WERE HIGHER AND ISSUED AMENDED REPORTS. THE LAB WAS USING THE AUTOMATED WEEKLY FLOW CELL CLEANING PROCEDURE. THE CUSTOMER PROVIDED SIX EXAMPLES OF THE ERRONEOUSLY LOW RESULTS ALONG WITH THE CORRECTED RESULTS. THE TOTAL NUMBER OF PATIENT SAMPLE RESULTS REPORTED OUT OF THE LABORATORY IS UNKNOWN. WHILE THE CUSTOMER DID NOT RECEIVE ANY REPORT OF ANY ADVERSE EVENT OR PATIENT INJURY ASSOCIATED WITH THIS EVENT, IT IS NOT KNOWN IF THERE WAS ANY CHANGE TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC SYNCHRON CLINICAL SYSTEM | JGS | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |