PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2011-82115
- Event Type
- Injury
- Date Received
- July 11, 2011
- Date of Event
- June 26, 2011
- Report Date
- June 27, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
THE CUSTOMER REPORTED BEING HOSPITALIZED DUE HYPERGLYCEMIA. THE BLOOD GLUCOSE READING WAS OVER 600 MG/DL. IT WAS STATED THAT THE CUSTOMER WAS TREATED AT THE HOSPITAL WITH AN INSULIN DRIP AND MANUAL INJECTIONS. THE CUSTOMER ALSO REPORTED HAVING A HEMATOMA TWO WEEKS AGO UNDER HIS SITE, AND AFTER REMOVAL HE WAS TAKING ANTIBIOTICS. THE CUSTOMER STATED THAT WHEN HE HAD HIGH BLOOD GLUCOSE, HE SAW THAT THE CANNULA WAS BENT. TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMING, TIME, AND DATE ON THE INSULIN PUMP WERE CORRECT. RAN A FIXED PRIME TEST AND PASSED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-522NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR | Hospitalization | QUICK-SET, 23| PARADIGM, 6MM CATHETER| (B)(4) INFUSION SET: MMT-399 |