FDA Adverse Event Malfunction Summary report: N

RAPIDPOINT 400

MDR report key: 217100 · Received March 30, 1999

Report

Report Number
1217157-1999-00001
Event Type
Malfunction
Date Received
March 30, 1999
Date of Event
March 1, 1999
Report Date
March 1, 1999
Manufacturer
CHIRON DIAGNOSTICS CORP.
Product Code
CCC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Description of Event or Problem · 1

CUSTOMER REPORTED UNDER-RECOVERY OF RAPIDQC COMPLETE QUALITY CONTROL MATERIAL FOR PARTIAL OXYGEN PRESSURE RESULTS. LEVEL 1 : ASSIGNED RANGE 134-156 MMHG; VALUES OBTAINED 126/127/130 MMHG; LEVEL 2: ASSIGNED RANGE 88-104; VALUES OBTAINED 83/85/97; LEVEL 3: ASSIGNED RANGE 12-34; VALUES OBTAINED 21/22. THIS INDICATED THAT THERE WAS POTENTIAL FOR UNDER-REPORTING PT VALUES FOR PARTIAL PRESSURE OF OXYGEN. IT IS BELIEVED THAT NO PT DATA WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RAPIDPOINT 400 BLOOD GAS ANALYZER CCC CHIRON DIAGNOSTICS CORP. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other