FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2170987 · Received June 30, 2011

Report

Report Number
1720753-2011-08660
Event Type
Malfunction
Date Received
June 30, 2011
Date of Event
June 22, 2011
Report Date
June 30, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED ON ONSITE INVESTIGATION. THE SVC REP REPLACED THE BATTERY CHARGER PRINTED CIRCUIT BOARD, AND CHECKED THE VOLTAGE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD DISPLAY A FAILED CHARGER ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1