COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-13297
- Event Type
- Malfunction
- Date Received
- July 21, 2011
- Date of Event
- June 29, 2011
- Report Date
- June 30, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP THAT EXPERIENCED FAILURE CODE 501:320:654:0000 WAS NOT CONFIRMED NOR REPRODUCED DURING PRODUCT EVALUATION; HOWEVER, THE QUALITY ENGINEER IDENTIFIED FAILURE CODE 199 TO BE THE CONFIRMED CONDITION. THE ROOT CAUSE OF THIS CONDITION WAS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES WERE REPLACED IN ORDER TO CORRECT THIS CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH THE REPORTED CONDITION OF FAILURE CODE 501:320:654:0000. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION IS UNKNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |