FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2170953 · Received July 21, 2011

Report

Report Number
2124215-2011-09989
Event Type
Malfunction
Date Received
July 21, 2011
Date of Event
June 6, 2011
Report Date
March 27, 2012
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

TO DATE, INFORMATION SUGGESTS THAT THIS MEDICAL DEICE REMAINS ACTIVELY IN SERVICE. AS NEW INFORMATION WOULD BECOME AVAILABLE, THIS REPORT WILL BE FURTHER EVALUATED AND UPDATED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THIS DEVICE WAS EXPLANTED FOR AN UNKNOWN REASON AND RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), WHICH IS PART OF THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS PRODUCT UPDATE CLASSIFIED AS A PRODUCT ADVISORY AND INITIALLY COMMUNICATED ON (B)(6) 2007, REACHED A MONITORING VOLTAGE MEASUREMENT OF 2.57 VOLTS AND A CHARGE TIME MEASUREMENT OF 13.8 SECONDS. THE BOSTON SCIENTIFIC TECHNICAL SERVICES CONSULTANT DISCUSSED THAT THESE MEASUREMENTS ARE WITHIN THE EXTENDED ERI CHARGE TIME LIMIT WINDOW BASED ON PROJECTED THERAPY USAGE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS CLINICAL OBSERVATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H217

Patients

Seq Age Sex Outcome Treatment
1 67 YR (B)(4)| MISMATCH| (B)(4)