HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-09280
- Event Type
- Malfunction
- Date Received
- July 21, 2011
- Date of Event
- June 24, 2011
- Report Date
- June 24, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A LOW DRAIN VOLUME ALARM WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. HOWEVER, BASED ON INVESTIGATION INFORMATION, THE CAUSE WAS DETERMINED AS AIR IN THE LINE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. ADDITIONAL INVESTIGATION IS BEING CONDUCTED THROUGH (B)(4).
(B)(4). THE SAMPLE WAS DISCARDED AND THE LOT NUMBER IS UNKNOWN. SINCE THE LOT NUMBER IS UNKNOWN, NO BATCH REVIEW WILL BE PERFORMED A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED.
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT A LOW DRAIN VOLUME ALARM OCCURRED ON HOME CHOICE (HC) DURING USE INITIAL DRAIN. THE NURSE(RN) STATED THAT THERE WAS AIR IN THE LINE. THE BAXTER TECHNICAL REPRESENTATIVE (TSR) EXPLAINED THAT THE RN WILL NEED TO END THERAPY AND START OVER WITH NEW SUPPLIES, MAKING SURE THAT THE PATIENT LINE WAS FULLY PRIMED BEFORE CONNECTING THE HOME PATIENT (HP) . THE TSR THEN WALKED THE RN THROUGH ENDING THERAPY PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | HOMECHOICE |