ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2011-10300
- Event Type
- Injury
- Date Received
- July 21, 2011
- Date of Event
- May 25, 2004
- Report Date
- July 14, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
AT THAT TIME AN INVASIVE EXAMINATION OF THE DEVICE SYSTEM WAS UNREMARKABLE FOR ANY ABNORMALITIES. THE RATE/SENSE IMPEDANCE HAD NORMALIZED. HIGH-ENERGY TESTING CONFIRMED APPROPRIATE DEVICE DETECTION AND THERAPY-DELIVERY, AS WELL AS LEAD IMPEDANCE MEASUREMENTS. NO CHANGES TO THE SYSTEM WERE MADE. ADDITIONAL INFORMATION RECEIVED INDICATES THIS LEAD WAS ATTEMPTED TO BE EXTRACTED DURING A DEVICE CHANGEOUT PROCEDURE. HOWEVER, THE EXTRACTION FAILED AND A PORTION OF THE LEAD PLANNED TO BE RETURNED. AGAIN, IT WAS NOTED THAT THE PACING IMPEDANCES WERE OUT OF RANGE. AT THIS TIME THE LEAD PORTION HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC, CRM FOR ANALYSIS. THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF FURTHER INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE RETURNED LEAD SEGMENT WAS PERFORMED. THE LEAD RETURNED WAS SEVERED 131 MM FROM THE TERMINAL PIN AND ONLY THE PROXIMAL SECTION WAS RETURNED. SET SCREW MARKS WERE NOTED ON ALL TERMINALS, 2 ON EACH. DRIED BLOOD/BODY FLUID WAS ALSO NOTED IN THE RS- LUMEN. ELECTRICALLY, THE LEAD SEGMENT RETURNED WAS CONTINUOUS. ANALYSIS WAS UNABLE TO CONFIRM THE REPORTED CLINICAL OBSERVATION WITH THE LEAD SEGMENT RETURNED.
GUIDANT RECEIVED INFORMATION APPROXIMATELY (B)(6) AGO THAT THIS IMPLANTABLE TRANSVENOUS DEFIBRILLATION LEAD MEASURED HIGH PACING THRESHOLDS AND AN OUT-OF-RANGE PACING IMPEDANCE OF >3000 OHMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Life Threatening| R | 0157| 1860| 0062| 1600 |