FDA Adverse Event Injury Summary report: N

FOCAL ONE

MDR report key: 21707622 · Received March 27, 2025

Report

Report Number
9612448-2025-00001
Event Type
Injury
Date Received
March 27, 2025
Date of Event
February 27, 2025
Manufacturer
EDAP TMS FRANCE
Product Code
PLP
PMA / PMN Number
K172721
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONCLUSION OF EDAP TMS INVESTIGATIONS SHOW THAT THIS FISTULA IS NOT RELATED TO A DYSFUNCTION OF THE DEVICE AND/OR THE PROBE. THE RECOMMENDATIONS OF THE USER MANUAL WERE NOT RESPECTED. IN FACT, WE HAVE AN INFILTRATING RECTAL WALL WITH A FAT LAYER THAT LARGELY REACTS TO THE TREATMENT. THIS RESULTS IN THE THICKENING OF THE RECTAL WALL AND OF A HYPOECHOGENIC AREA (FAT LAYER) EXCEEDING 1 CM AND WHICH SEEMS DIRECTLY RELATED TO THE CREATION OF THE URETHRORECTAL FISTULA. IN ADDITION, THE TREATMENT WAS NOT STOPPED EVEN WITH A PERSISTENT FOGGY AND THEN BARRIER CAVITATION LEADING TO A BLIND TREATMENT (IMPOSSIBLE TO SEE THE PROSTATE CONTOUR, TRAJECTORIES OUTSIDE THE PROSTATE CAPSULA AND FIRST FOCAL POINTS ON THE RECTAL WALL). NOTE THAT THIS TYPE OF CAVITATION ALREADY APPEARS DURING THE TREATMENT OF THE FIRST AREA (RIGHT LOBE).

Description of Event or Problem · 0

THERE HAS BEEN A REPORTED FISTULA FROM (B)(6). DOCTOR SENT THE FOLLOWING TEXT MESSAGE TO EDAP TMS (2/27/2025): "UNFORTUNATELY HAVE A PROSTAORECTAL FISTULA ON A RECENT HIFU ((B)(6)) DONE ON (B)(6) 2025".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1438525 FOCAL ONE FOCAL ONE PLP EDAP TMS FRANCE 239320

Patients

Seq Age Sex Outcome Treatment
1 NA Male Other