FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2170741 · Received June 21, 2011

Report

Report Number
3004209178-2011-81874
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
June 4, 2011
Report Date
June 4, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY AS A RESULT OF MOISTURE DAMAGE FOUND ON THE ELECTRONIC AND MOTOR ASSEMBLY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP FELL IN WATER. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT SHE REMOVED THE BATTERY AND WATER INSIDE OF THE INSULIN PUMP WAS FOUND. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAL

Patients

Seq Age Sex Outcome Treatment
1 26 YR