FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2170741
·
Received June 21, 2011
Report
- Report Number
- 3004209178-2011-81874
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Date of Event
- June 4, 2011
- Report Date
- June 4, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY AS A RESULT OF MOISTURE DAMAGE FOUND ON THE ELECTRONIC AND MOTOR ASSEMBLY.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP FELL IN WATER. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT SHE REMOVED THE BATTERY AND WATER INSIDE OF THE INSULIN PUMP WAS FOUND. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-522NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR |