FDA Adverse Event Malfunction Summary report: N

ENDOPATH** ETS45

MDR report key: 2170602 · Received July 21, 2011

Report

Report Number
3005075853-2011-02930
Event Type
Malfunction
Date Received
July 21, 2011
Date of Event
June 24, 2011
Report Date
June 24, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE (B)(4) DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES NOTED. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC RIGHT HEMICOLECTOMY PROCEDURE, THE DEVICE WAS LOADED CORRECTLY WITH A VASCULAR RELOAD AND HANDED TO THE SURGEON. THE SURGEON FIRED THE DEVICE ACROSS THE RIGHT ILEO COLIC ARTERY BUT THE VESSEL BLED IMMEDIATELY FOLLOWING THE JAW RELEASE. IT APPEARED FROM THE CARTRIDGE THAT A ROW OF STAPLES MAY NOT HAVE BEEN FIRED. THE SURGEON HAD TO OVERSEW THE STAPLE LINE. AS A RESULT OF THE DIFFICULTIES ENCOUNTERED WITH THIS DEVICE, THE PROCEDURE WAS PROLONGED TEN MINUTES. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** ETS45 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK H43K10

Patients

Seq Age Sex Outcome Treatment
1