ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2011-05093
- Event Type
- Malfunction
- Date Received
- July 21, 2011
- Report Date
- June 20, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE CARTRIDGE WAS NOT RETURNED, A RETAINED SAMPLE WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE CARTRIDGE PASSED VISUAL INSPECTION, NO DAMAGE OR DEFECTS WERE NOTED TO THE LUER CONNECTION OR O-RINGS. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE.
LOT #: B201636, SERIAL #: N/A. THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PATIENT REPORTED LEAKY CARTRIDGES FROM LOT # B201636. HE STATED THAT IT HAS BEEN OCCURRING FOR ABOUT A MONTH. THE PATIENT REPORTED THAT THE CARTRIDGES SEEM TO BE LEAKING FROM THE BOTTOM; HE HAS NOTED AIR BUBBLES IN THE CARTRIDGES AND THE TUBING. HE STATED THAT HE HAS TO RE-PRIME TO CLEAR THE AIR BUBBLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN PUMP CARTRIDGE | LZG | ANIMAS CORPORATION | ANM IR1200/1250/2020/OTP CART |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |