COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-13263
- Event Type
- Malfunction
- Date Received
- July 21, 2011
- Date of Event
- July 1, 2011
- Report Date
- July 1, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY WAS CONFIRMED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM WAS DETERMINED TO BE DEPLETED BATTERIES RESULTING FROM USER ERROR. APPROPRIATE ACTION WAS TAKEN TO CORRECT THE REPORTED PROBLEM BY REPLACING THE MAIN BATTERIES AND HARNESS. A SERVICE HISTORY REVIEW FOUND THAT THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION. HOWEVER, DURING PREVIOUS SERVICE THE BATTERIES AND HARNESS WERE REPLACED. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.
(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE PUMP BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. THIS EVENT OCCURRED UPON POWER UP; HOWEVER IT IS UNKNOWN IN WHICH CARE AREA THIS EVENT OCCURRED. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |